Leica IP C Instructions For Use Manual page 54

Automated printing system for tissue cassettes
Hide thumbs Also See for IP C:
Table of Contents

Advertisement

5
Operation
• Once all trays are full the instrument emits an acoustic signal (beep), the LED in the TRAY LOAD
button starts blinking, the stack of trays can be removed.
Fig. 44
Controls the movement of the lifting table of the automated unload station (optional)
1. Place a stack of trays
2. Any number of trays between 1 and 10 can be inserted, as the printer counts the trays when
they are inserted.
3. Once the lifting table has reached its upper limit position, the LED
starts blinking.
4. Press and hold TRAY LOAD longer than 1 sec.
5. The tray stack moves completely into the unload station, the LED in the button goes off, the
printer switches to ONLINE mode.
6. Pending print jobs will be carried out.
7. When the stack of trays has moved
8. Press TRAY LOAD
9. The stack of trays moves up by one tray.
If TRAY LOAD is pressed and held longer than one second:
» The stack of trays moves completely out of the unload station, the LED in the button starts
flashing. Any print job in progress is interrupted.
Note
Every time the printer is switched on, the stack of trays automatically moves one tray up, to ensure
that the new print job is started with an empty tray.
Warning
Be careful about getting near the sensor
sensor will trigger a lifting movement.
54
1
(→ "Fig. 43-1")
onto the lifting table of the unload station
completely or partly
briefly:
(→ "Fig.
(→ "Fig. 42-9")
into the unload station:
44-1"). Any object getting closer than 2 mm to the
(→ "Fig.
43").
in the button
Version 2.0, Revision I