Installing/Exchanging The Flash Bulb - Leica IP S Instructions For Use Manual

Table of Contents

Advertisement

4.
Instrument Setup

4.7 Installing/exchanging the flash bulb

41
Fig. 16
43
28
49
42
Fig. 15
Caution!
To insert/remove, hold the flash
bulb as shown in
Do not take hold of the sides of
the flash bulb or compress it as
shown in
Fig.
44
Fig. 18
Removing the old flash bulb
Switch the instrument off and unplug it
from power supply.
Allow the flash tube to cool before re-
moving it. Do not handle the flash tube
with bare hands. Use a glove or tissue.
• Open the lid (4,
reflector.
• Remove screw (49) (use screwdriver supplied
as part of tool set). Mind the washer (42).
Fig. 17
(right).
16.
• Swing the reflector (41) upwards.
• Carefully pull out the old flash bulb (43) straight
to the right, do not twist it. If the flash bulb
cannot be pulled out easily, gently rock it back
and forth to get it unseated from the socket.
• Make sure the contact spring (44) is removed
from the priming wire (45) of the lamp.
(See also
Fig. 19
Instructions for Use V 1.9 RevD - 05/2013
Fig.
2) to gain access to the
and 22)
Fig. 17