7.

Troubleshooting

7.1

Malfunctions

Note
If the printer malfunctions during printing, a corresponding error code is displayed and
simultaneously the LED in the ERROR button starts flashing.
How to eliminate the problem:
1. Determine the cause of the error using the error list
2. Remove obstruction(s); open lid if required.
Warning
Important!
Remove all cassettes which are still in the chute, in or next to the cassette carrier or in the drying
module. These cassettes should not be reused.
Confirm elimination of error:
1. Close the lid and press ERROR to confirm to the printer that the source of the error has been
eliminated.
2. The printer then verifies whether all processing paths are unobstructed and whether the
source of the error has been eliminated.
3. If there are still some obstructions left, or if the source of error has not been
eliminated, the printer displays another error message.
4. Interrupted print jobs are resumed where previously left off.
5. If an error message is displayed several times although all possible causes have been
eliminated, a Reset should be carried out.
Leica IP C
Troubleshooting
(→ p. 75 – 7.3 Error
messages).
7