7
Troubleshooting
Display
code
A print job was attempted with a storage
fluid cartridge inserted.
Power supply unit does not achieve
standard operating voltage.
Firmware only partially loaded or not
loaded at all.
Flash memory defective.
Wrong firmware.
7.4

Changing the flash bulb

Code
13
appears on the printer display when the flash bulb has reached its maximum service life.
When code
is issued, the bulb has to be changed.
13
For details on how to insert/replace the flash bulb, please refer to
the flash
bulb).
Warning
Before replacing the flash bulb, switch off the printer and unplug the instrument from power supply.
10 s
1. After changing the bulb, switch the printer back on.
2. Then go to offline mode and hold down the ERROR key for 10 seconds; code "13" disappears
from the display.
78
Source of error
Troubleshooting solution
Remove storage cartridge.
cartridge and press LOADED to confirm
(→ p. 41 – 4.9 Exchanging the transport
cartridge for an ink
cartridge).
Call Leica Technical Service.
Call Leica Technical Service.
Call Leica Technical Service.
(→ p. 32 – 4.6 Installing/exchanging
Version 2.0, Revision I