Display
code
Cassette in the drying module during
initialization or prior to a print job being
carried out.
The flap of the drying module either does
not close or does not open. The flap is
blocked (e.g. by a cassette).
Lifting table end sensor does not switch.
Lifting table position sensor does not
switch during positioning.
Flash power supply: Overtemperature for
more than 10 minutes.
Faulty control data received (program bug).
Settings for the serial interface incorrect or
the instrument configuration conflicts with
the PC configuration.
Transmitted data do not contain
confirmation of receipt or data
transmission was not confirmed by the PC.
Print image exceeding the vertical limit.
Print image exceeding the horizontal limit.
The CRC test of the EEPROM returned an
error when the instrument was switched
on.
Internal firmware error or defective
controller.
Mechanical problem makes ejection of
cassettes from magazine difficult.
Leica IP C
Source of error
Troubleshooting
Troubleshooting solution
Remove the cause of the blockage.
Remove the cause of the blockage.
Call Leica Technical Service.
Call Leica Technical Service.
Call Leica Technical Service.
Carry out a RESET on the printer.
Check cable connection to the PC.
Check configuration of serial port of the
PC and reboot the PC.
Follow the same procedure as for
ERROR
60.
Try another printer cable.
Error caused by application software.
Error caused by application software.
Call Leica Technical Service.
Call Leica Technical Service.
Check ejection mechanism. Remove
foreign bodies, then clean with brush.
7
77