Transporting The Instrument - Leica IP C Instructions For Use Manual

Automated printing system for tissue cassettes
Hide thumbs Also See for IP C:
Table of Contents

Advertisement

6
Cleaning and Maintenance
Warning
• Switch off the printer and disconnect the power to prevent damage to the print head!
• Never use a transport cartridge together with a
• In order to use the printer again, the transport plate has to be removed and a new exchange plate
has to be inserted.
• Do not reinstall a used transport plate, as it will no longer completely seal the print head.
22. Likewise, clean the removed
(→ "Fig. 63-2")
23. The sealing lip
damage.
Warning
Do not reuse a
replacement plate
3
Fig. 63
24. Clean the instrument completely, as described in this chapter.
6.6
Transporting the
If the printer is to be shipped, the storage instructions described above are to be carried out completely.
Then, additionally make the following preparations:
1. Open the hood
2. Then take out the flash bulb. For more information, refer to
exchanging the flash
3. Close the flash cover and the hood.
4. Use original packaging and securely screw the printer to the baseplate (see unpacking
instructions).
5. Reinsert the transport anchor
6. Make sure that the instrument is only transported upright.
72
replacement plate
with (clean) alcohol and a cleaning swab
(→ "Fig. 63-2")
must be completely free of ink residue. Check the sealing lip for
with a damaged sealing lip!
1
2
instrument
(→ "Fig. 34-1")
of the printer and loosen the screw of the flash cover.
bulb).
(→ "Fig. 14-1")
replacement
plate!
(→ "Fig. 63-1")
with the sealing lip
(→ "Fig.
63-3").
(→ p. 32 – 4.6 Installing/
and secure the hood with adhesive tape.