At the top are the elements detected in the sample, and underneath this,
elements that are below the detection limit.
Use the 4 way touch pad or touch screen to scroll through the elements.
For an element to be detected by your analyzer in a given sample, the
measured concentration of the sample must be at least three times the
standard deviation of the measurement. This detection limit will depend on
the composition of the sample.
Precision
The measurement precision for each element displayed appears to the right
of the measured concentration, under the heading "+-". The precision of
each measurement is two times the standard deviation (sigma).
An element is classified as "detected" if the measured concentration (in
ppm) is at least 1.5 times the precision. Detected elements are displayed in
ppm, followed by the measurement precision. Non-detected elements are
shown as < the detection limit for that sample. The detection limit for a
given element varies depending on the other elements in the matrix.