Analyzing Bulk Samples
WARNING! The preconditions for operation must be continued for the
duration of the reading. If the preconditions are violated, the x-ray tube will
turn off, the calibration shutter will close, and the measurement will end.
The four LED lights will stop blinking when the measurement is ended.
The flashing of the LED lights is not synchronized to minimize power
consumption.
To end the test, simply release the trigger mechanism, or any other
applicable preconditions.
WARNING! When all four LED lights are blinking, the x-ray tube is on.
This should only occur during a measurement, while the preconditions for
operation are met. On startup, the front pair of lights will blink. If the LED
lights blink at any other time, disconnect the battery pack and call Thermo
Scientific's Service Department in the United States, toll free, at (800)
875-1578, or outside the United States, at +1-978-670-7460, or your local
Authorized NITON Analyzer Service Center.
Your NITON Analyzer will display the Results Screen throughout the
duration of each reading, The Results Screen is updated regularly
throughout the reading. When the reading is complete, a final screen update
will appear, and your NITON analyzer will display the final results of the
measurement which has just been completed.
WARNING! Do not attempt to take measurements while downloading
readings! This will generate an error requiring a system reset, and may
corrupt your stored readings, requiring all stored readings to be erased.
Thermo Scientific
NITON XL3 Analyzer User's Guide
1-11