Calibrating the Touch Screen Without Using the Touch Screen
1-66
NITON XL3 Analyzer User's Guide
5. Move the cursor to the appropriate first number in your
password and then using the enter button on the right (it has the
arrow/enter key symbol on it) press this "enter" key. The first
number of your password should appear in the lower left of the
LCD screen.
6. Repeat step 5 until you have entered the entire password. Then
move the cursor to the letter "E" and press the "enter" key to
enter it.
7. You will now be at the main screen.
8. Again using the four way touch pad, move the cursor to highlight
the "Utilities" icon and press the "enter" key to select.
9. You will now be at the Utilities screen
10. From the Utilities screen, move the cursor to highlight the
"Calibrate" icon and press the "enter" key.
11. Now move the cursor to highlight "Calibrate Touch Screen" and
press the enter key.
12. You are now at the Touch Screen Calibration screen.
Note You must now use the touch screen for the balance of this procedure
13. In the upper left hand corner you will see a crosshair "+", using
the stylus or a pen, tap the center of the "+".
14. Repeat this for each "+" sign that appears, there should be one
for each of the 4 corners.
Your touch screen should work properly after this and you may use it from
this point forward. If it does not, please repeat the process.