Atria Electrocardiograph Operating Instructions
Electromagnetic interference. Position the Atria away from other electrical
or electronic equipment, if possible. The presence of strong EMI fields, or
generated by RF noise on the line power, or by electronic, surgical, or
diathermy instruments in close proximity to the Atria may cause trace
noise or input overload conditions.
If used with, or around, other electrical or electronic equipment, always
carefully monitor initial readings to verify normal operation.
The Atria is compliant with IEC 60601-1-2 EMC immunity requirements.
Refer to "EMC Declaration Tables" below.
The Atria is intended for use in the electromagnetic environment specified
below.
Electromagnetic Emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
WARNING
Electromagnetic environment -
Compliance
Group 1
The Atria uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Class B
The Atria is suitable for use in all
establishments.
Class B
Complies
guidance