Communications Setup; Chapter 5 Communications Setup - Cardiac Science Atria 6100 Operating Instructions Manual

Electrocardiograph
Table of Contents

Advertisement

5
Chapter
!
70-00082-01 A

Communications Setup

Misdiagnosis. Electromagnetic interference may cause trace noise or
input overload conditions. This may be caused by the presence of strong
EMI fields, or generated by RF noise on the line power, or by electronic,
surgical, or diathermy instruments in close proximity to the Atria. Position
the Atria away from other electrical or electronic equipment, if possible.
If used with, or around, other electrical or electronic equipment, always
carefully monitor initial readings to verify normal operation.
The Atria is compliant with IEC 60601-1-2 EMC immunity requirements.
Refer to "EMC Declaration Tables" on pg. 11-10.
External communications are optional on the Atria.
For assistance in determining which options are available on your Atria, or
to purchase additional options, call Customer Service at (800) 777-1777 or
(608) 764-1919.
Base communications options include:
• Serial Communications (RS232)
• Wired Ethernet
• USB
• Analog Out
The following features require the Communications options:
• Email
• Fax
• Plain paper printing
• EMR communications
Additional communication options include:
• Internal Fax/Modem
• BlueTooth connection
• 802.11 network connection
• PPP connection
NOTE: Wireless options transmit at 2.4 GHz. Other
nearby wireless devices may cause interference. If possible,
move or turn off other devices to minimize potential
interference.
Setting up any of the communications options listed above may require
assistance from the Network Administrator or IT department.
WARNING
Atria Electrocardiograph Operating Instructions
5-1