Acquiring a Stat ECG

Recommended Workflow

!
70-00082-01 A

Acquiring ECG Reports

The Atria does not require patient demographic information for an
immediate ECG. Enter patient demographic information at any time. (See
"ECG Tips" on pg. 7-4.)
If no patient ID is entered, the Atria prints "#STAT#, date, time" in the
field.
Patient ID
To acquire a stat ECG:
1. Prepare the patient for the recording. See "Patient Preparation" on
pg. 6-1.
2. Press [On/Stby] if necessary.
3. Verify the patient's heart rate displays on the
does not display the heart rate, check the problem leads listed on the
display. Replace faulty electrodes if necessary.
4. Press [ECG] twice to acquire a stat ECG report.
NOTE: Pressing [ECG] twice overrides the "Wait for Good
Data" message and may result in a flat-line ECG because the
Atria hasn't acquired enough good data.
If possible enter patient demographic information before performing tests.
Patient demographic information also may be entered after acquisition.
Misdiagnosis. Some patient information directly affects ECG analysis.
Your patient's physician uses this information when interpreting ECG
reports. Likewise, an interpretive Atria provides more accurate and
complete analysis statements when you enter complete patient
information.
Prepare the Patient
Prepare the patient for the recording, see Chapter 6, "Patient Preparation".
Home
WARNING
Atria Electrocardiograph Operating Instructions 7-1