Guidance and manufacturer's declaration—electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Powerheart® AED G3 Plus 9390A and 9390E
IEC 60601 test level
Compliance level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power supply
Not applicable
lines
±1 kV for input/output
lines
±1 kV differential mode
Not applicable
±2 kV common mode
Electromagnetic emissions standards compliance
Electromagnetic environment—
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
70-00914-01 F
1-13