11
Maintenance and Service

Technical Specifications

Dimensions
Weight (unit only)
Display
Keyboard
Data Storage
Input Power
Battery duration
Printout
Printout device
Paper dimension
Paper type
Chart speeds
Gain
Printout formats
Acquisition
Lead selection
Modes
Frequency response
Input impedance
Electrode offset tolerance
Sampling
Artifact filter response
Storage
Pacemaker display
capability
Interpretation (optional)
11-14
Atria Electrocardiograph Operating Instructions
15.0" x 13.125" x 5.5"
(381mm x 334mm x 140mm)
11 lbs (5 kg) (including external power supply)
Atria 3100: 2 x 40 character backlit LCD screen
Atria 6100: 640 x 480 backlit color LCD screen
Full alphanumeric keypad plus designated quick keys
Atria 3100: 50 records standard
Atria 6100: 150 records standard, optional upgrade to 300
AC Operation 100-240 VAC ± 10%, 50-60 Hz ± 3 Hz
55 minutes continuous printing
216 mm thermal dot array
8.5" x 11" (US letter)
210mm x 300mm (A4)
Thermal sensitive (Burdick Assurance
paper recommended)
12.5, 25, 50 mm/sec
5, 10, 20 mm/mV Chest or Limb (may be split)
3, 4, 6 or 12 channels; additional rhythm formats
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 and
Alternate Chest Lead (chest lead selection V2R through
V9R, V7, V8, and V9)
Automatic, automatic rhythm, or manual rhythm
Meets or exceeds ANSI/AAMI EC11-1991 and IEC 60601-
2-51 standards
Meets or exceeds ANSI/AAMI EC11-1991 and IEC 60601-
2-51 standards
±300 mV
8000 Samples/Sec. 2.5µV LSB
40 Hz, -3db
500 samples/second, 2.5µV LSB
Meets or exceeds ANSI/AAMI EC11-1991 standard and
IEC 60601-2-51
Diagnosis, measurements, reasons statements based on
five demographic criteria
®
or HeartLine™
70-00082-01 A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Atria 3100

Table of Contents