Measurement Matrix

Atria models with interpretation and measurement capabilities also have
the Measurement Matrix printing option. This option is available after
acquiring the ECG via the
Patient Record" on pg. 8-2 to print the measurement matrix using a saved
patient record).
The Measurement Matrix consists of a column for each standard lead: I, II,
III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
The following table explains the numerical values in the Measurement
Matrix.
NOTE: Unless otherwise specified, all times are in
milliseconds and all amplitudes are in microvolts.
Units
ms
Time from the beginning of the representative beat to the beginning
of the P wave.
ms
P wave duration (total).
ms
Time from the beginning of the representative beat to the beginning
of the QRS complex.
ms
QRS duration.
ms
Q wave duration.
ms
R wave duration.
ms
S wave duration.
ms
R' wave duration.
ms
S' wave duration.
ms
P+ wave duration.
ms
Intrinsicoid deflection time.
µV
P+ wave amplitude.
µV
P- wave amplitude.
µV
Peak to peak amplitude of the QRS complex.
µV
Q wave amplitude.
µV
R wave amplitude.
menu (see also "How to Display a
Patient Review
DESCRIPTION
Atria Electrocardiograph Operating Instructions