Guidance and manufacturer's
declaration—electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified
below. The customer or the user of the AED should assure that it is used in
such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Version 112-2025-109 B
IEC 60601 test
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common mode
Compliance level
±6 kV contact
±8 kV air
Not applicable
Not applicable
Product Information and Safety 1-13
Electromagnetic environment—
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%