Test Unit Address Allocation - Sartorius Sartocheck 4 plus Operating Instructions Manual

Bag tester / bag tester multiunit
Hide thumbs Also See for Sartocheck 4 plus:
Table of Contents

Advertisement

Available languages

Available languages

7.19.1.3 Test unit address
Fig. 7-70 | Service Menu: Hardware configuration
Fig. 7-71 | Service Menu: Hardware configuration
This function sets the address of the test unit so that the Sartocheck
can allocation recognize it. The address Test Unit #1 is reserved for the Sartocheck
plus Bag tester.
Allowable addresses for the MultiUnits are Test Unit #2 through Test Unit #5.
MultiUnits come from the factory with no address allocation, and therefore
must be set by the user.
When connecting MultiUnits to the Sartocheck
is important that only one new MultiUnit be connected at a time. Two or more units
that have the same address allocation, or no address allocation (default setting from
the factory), will cause a communication conflict.
When transferring a MultiUnit to be used with a different Sartocheck
tester, it is advisable to first set the address allocation back to the default settings.
MultiUnit address allocation procedure:
t
MultiUnit should be turned on, but the communication cable to the MultiUnit
should not be connected.
t
From the Hardware configuration screen of the Sartocheck
press F3 (Test Unit address allocation).
t
Connect the MultiUnit communication cable when prompted by the
®
Sartocheck
4 plus Bag tester and press "Page down".
t
The new MultiUnit will be displayed in F7 as "New test unit" with the appropriate
serial number.
t
Select F7 and press "Page down".
t
Available test unit address locations will be shown.
t
Select the desired address location and press "Page down".
Change a MultiUnits address allocation:
t
From the Hardware configuration screen of the Sartocheck
press F3 (Test Unit address allocation), then press "Page down".
t
Select the appropriate test unit and press "Page down".
t
Select the desired address location and press "Page down".
To set the MultiUnit back to the factory default setting, select F7
[Reset to standard address] and press "Page down".
®
4 plus Bag tester
®
4 plus Bag tester for the first time, it
®
4 plus Bag
®
4 plus Bag tester,
®
4 plus Bag tester,
Operation
®
4
57