4.3 Installation of

4.4 Shutdown Procedure

14
General Warnings
Please note that replacement parts and Replacement Parts accessories that not have
been supplied by Sartorius Stedim Biotech GmbH have not been tested and approved
by Sartorius Stedim Biotech GmbH. The installation of and |or the use of such
products can therefore have a negative effect on the design properties of the unit as
constructed.
Sartorius Stedim Biotech GmbH is not liable for any damages resulting from the use
of non-original parts or non-original accessory parts.
Standard parts can be purchased from specialized dealers.
The following shutdown procedure must be followed before any cleaning,
servicing or repair work is carried out.
1. Line disconnect
Turn off the main switch by pulling out the AC adapter of the mains supply.
2. Compressed air supply
Turn off the compressed air supply and vent system.
Personnel may suffer physical injuries or fatalities if this procedure is not followed!