General Information
1.7
Integrity Tests and Integrity Test Data
After wetting, Maxicaps
recommended that testing is performed by an automatic integrity testing
device like Sartocheck
For the Maxicaps
®
Test pressure:
Stabilization time:
Testing time:
Alarm values for the corresponding product:
3VI--28-MCG-MR2
3VI--28-MCG-MR3
3VI--28-MCG-MR4
3VI--28-MCG-MR5
3VI--28-MCG-MR6
3V2--28-IVG-MR3
3V2--28-IVG-MR6
For the Maxicaps
®
Test pressure:
Stabilization time:
Testing time:
8
Instructions for Use Maxicaps
MR Virosart
®
.
®
MR Virosart
HF and Media the test conditons are:
®
2.5 bar | 36 psi
5 min
5 min
MR Virosart
Max the test conditons are:
®
2.0 bar | 29 psi
5 min
5 min
can be integrity tested. It is highly
®
≤ 41.0 mL/min
≤ 60.0 mL/min
≤ 79.0 mL/min
≤ 99.0 mL/min
≤ 117.0 mL/min
≤ 48.0 mL/min
≤ 97.0 mL/min
MR Virosart
HF, Max and Media
®
®