Fig. 7-56 | Cleaning
Fig. 7-57 | Cleaning
Fig. 7-57b | Cleaning
Fig. 7-57c | Cleaning
Description of the connections:
a:
IN – Inlet pressure tubing
Connect inlet pressure tubing
b:
Venting 2
Connect this outlet as defined
c:
OUT – Outlet pressure tubing to the filter housing
Connect this outlet as defined
d:
Venting 1
Connect this outlet as defined
e:
Ext.Ref.Tank / Venting 3
Connect this outlet as defined.
For the Sartocheck 4 plus Bag tester you will need a coded Stäubli
connector, see accessories.
f:
Shut-off valve
This valve may only be opened once all the cleaning equipment
is installed and the cleaning procedure is started.
g:
Pressure Vessel
Connect the pressure vessel as defined and fill with cleaning fluid.
h:
Collecting Flask
Connect the collecting flask as defined. This vessel serves to collect the
cleaning | rinsing fluid (drain) used.
The cleaning fluid we recommend:
Water (cold, warm or hot; never use warm or hot water for protein contamination!)
Buffer solutions, e.g. 0.9 % NaCl Solution
Sodium hydroxide solution, max. 1 M
In step 5, clean all the pneumatic components, i.e. the pneumatic block and
the internal reference tank are flushed with water. This is done to ensure that no
detergents remain in Sartocheck
If you have not installed the connections in accordance with the diagram and the
procedure has already started, we recommend that you cancel cleaning and restart.
The pneumatic components can be damaged or their function impaired by improper
connection. Improperly connected tubing can also impair the effectiveness of the
cleaning procedure. Therefore, cleaning should be restarted.
The duration of the individual cleaning steps is predefined at 10 min per step.
If it is necessary to change the time for the cleaning and drying steps, this can be
done within the Service Menu.
To access this menu from the main menu, press F7 Service menu, F2 Software
configuration, F10 Test unit parameters, F3 Edit parameters, F1 Cleaning | Drying.
®
4 plus Bag tester.
Operation
51