14.5 Outlets

82
PLC Control
Name
Function
Outlet D
Unit is ready to use (stand by)
Outlet E
Test running
Outlet F
Error display
Outlet G
Test passed (OK)
Outlet H
Test failed (not OK)
Outlet L
Ext. valve 1
Outlet M
Ext. valve 2
Note 1: Outlet L not activated during a test until the end of stabilization time
Note 2: Outlet M not activated during a test until the end of measuring time
Plug Layout
Plug
Function
Pin A
START
Pin B
STOP
Pin K
PROG
Pin C
GND
Pin D
POWER ON
Pin E
TEST RUNNING
Pin F
ERROR
Pin G
TEST OK
Pin H
TEST NOT OK
Pin J
+24 V DC
Pin L
EXT. VALVE 1
Pin M
EXT. VALVE 2
Outlet Status
D
E
F
G
H
1
0
0
0
0
1
1
0
0
0
1
1
1
0
0
1
1
0
1
0
1
1
0
0
1
see note 1
see note 2
L
M
0
0
1
2
1
1
0
0
0
0
0
0