7.13 Plate Holder Test

Fig. 7-27 | Test parameters for plate holder test
Fig. 7-28 | Data log group 1
Fig. 7-29 | Data log group 2
40
Operation
This test sequence is designed for leak testing of 2 dimensional bags (e.g. Flexboy)
being maintained between 2 metal plates during the test e.g. using the FlexAct BT
trolley equipped with fleeces. The test sequence is based on the change in pressure
within the bag system including connected tubes.
The pressure drop measured in the process can be used as a direct measure of the
bag set-up leak tightness and absence of pinholes or similar defects, provided that
the upstream volume of the system is constant.
Enter parameters for the plate holder test
Test pressure:
from 3–300 mbar
Stabilization time:
10-3600 seconds.
Test time:
10-3600 seconds
Max. pressure drop:
1-programmed test pressure
Please verify the test parameters for the bag being tested. Applying too high a test
pressure may jeopardize the integrity of the bag.
Since the viscosity of the remaining liquid may influence the leak rate of an eventual
pinhole, please do not mix up pre-use test values with specifically validated post-use
values.
This test can also be used for testing vessels as long as they resist the applied
pressure.
Enter record data 1 (up to 32 characters per field)
Company:
Company name
Building:
Building name; house number
Department:
Identification of the department
Manufact. site:
Production sector
Product:
Product name
Product lot:
Batch number
Bag type:
E.g. Flexboy 50L
Bag lot:
Number of the bag batch
Locks
After entering with the Shift + F keys, the programmer can lock the data to prevent
any changes. The dialog boxes are then highlighted in gray. The user can no longer
change the entered data. The relevant record data are entered in the menu
"Test program: Record data 1."
Entry of record data 2
Nom. bag volume:
Nominal volume of the bag
Test gas:
Type of gas
Comment:
Name of operational process
Comment 2:
Enter test-relevant remarks
Comment 3:
Enter test-relevant remarks