System Settings
7.3.6
Parameters in the "Language" Main Menu
Parameters
Setting values
Language
ENGLISH*
DEUTSCH
FRANCAIS
ITALIANO
ESPANOL
PORTUG.
РYССKИИ
POLSKI
* Factory setting
7.3.7
Parameters in the "Unit 1", "Unit 2", "Unit 3", and "Unit 4" Application Menus
Parameters
Setting values
NONE*
GRAMS – NEWTON NONE*
ALL
LP .ON/OFF
DIVIS . 1
MINUS 1
* Factory setting
48
Cubis
®
MCE Operating Instructions
Explanation
English
German
French
Italian
Spanish
Portuguese
Russian
Polish
Explanation
The availability of units may depend on national legislation and is therefore country-
specific .
Switches the toggle between weight units off for this position .
The configuration values for the [DISP .DIG .] parameter in the [BALANCE] menu are
saved for the scale interval ([DISP .DIG .] parameter see Chapter 7 .3 .1, page 37) .
"Show all decimal places": All decimal places are shown in the display .
"Reduced by 1 decimal place for load change": The last decimal place on the display is
switched off until stability is achieved .
"Last decimal place of the 1st division": The last decimal place always shows the 1st
division .
"Last decimal place off": The last decimal place is switched off and the measured value
is rounded up or down by one decimal place .