Terminal Assignment - Endress+Hauser Proservo NMS5 Operating Instructions Manual

Hide thumbs Also See for Proservo NMS5:
Table of Contents

Advertisement

5 Wiring
34
5.2

Terminal Assignment

5.2.1
Ex d, Output 1 : V1, Modbus RS485, HART, Enraf BPM
V1, Modbus RS485, HART, Enraf BPM
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External contact Input
(Option)
DC4...20mA Output (Option)
DC4...20mA Output (Option)
NMT or HART
Device (NonIS)
or Pt100 ohm Input
Order Code Example
NMS5 -
Figure. 36: Electrical Connection of NMS5
14
16
18
20
22
CTR2
OT1+
OT2+
AL4 COM
AL4
N.C.
CTR1
OT1-
OT2-
15
17
19
21
AC Power Supply
Ground
Lightning Arrester
NRF or Non IS HART Devices
Common
Hoist
Spot
X
X
X
N
1
P
5
F
6
A
G
J
Q
B
S
C
B
D
C
G
E
H
Proservo NMS5
25
26
24
NMT+ NMT-
23
A
B
b
Fuse
F1
POWL
1
POWN
2
POWG
3
L
ARSL
L
ARSG
G
G
ARSN
N
N
+
NRF+
4
-
NRF-
5
+
RCA/+
6
-
RCB/-
7
ALM1/OSP
8
ALM1/OSP
9
ALM2/OSP
10
ALM2/OSP
11
ALM3
12
ALM3
13
ALM4
14
ALM4
15
COM
16
CTR1
17
CTR2
18
19
OUT1+
+
20
OUT1-
-
21
OUT2+
+
22
OUT2-
-
23
TMPA /NMT+
+
24
TMPB/NMT-
-
25
TMPb
26
X
X
X
X
X
X
Endress+Hauser
X
X

Hide quick links: