4.3 CALCulation Group
PARameter
POLarity
:CALCulation
Function
Queries all settings related to the statistical
value.
Syntax
:CALCulation?
Example
:CALCULATION? -> :CALCULATION:
AREA BLOCK;PARAMETER:AVERAGE 1;
DEVIATION 0;FLUTTER 0;JITTER 0;
MAXIMUM 0;MEDIAN 0;DEVT 0;
MINIMUM 0;MODE 0;PTOPEAK 1;
SDEVIATION 1;TAVERAGE 1;TFLUTTER 0;
TJITTER 0;TMAXIMUM 0;TMINIMUM 0;
TPTOPEAK 1;TRF 1;TSDEVIATION 1
4-10
;
:
CLEar
AVERage
<Space>
DEViation
FLUTter
JITTer
MAXimum
MEDian
DEVT
MINimum
MODE
PTOPeak
DEViation
TA Verage
<Space>
TFLutter
TJITter
TMAXimum
TMINimum
TPTopeak
TRF
TSDeviation
<Space>
POSitive
NEGative
BOTH
POSNeg
?
?
OFF
ON
<NRF>
?
OFF
ON
<NRF>
?
?
:CALCulation:AREA
Function
Queries the computation range of the statistical
values or queries the current setting.
Syntax
:CALCulation:AREA
{MARKer|WINDow|BLOCk}
:CALCulation:AREA?
Example
:CALCULATION:AREA MARKER
:CALCULATION:AREA? -> :CALCULATION:
AREA MARKER
Description • Cannot be specified when using multi window
or auto window.
• BLOCk cannot be specified when using
histogram.
• BLOCk cannot be specified when measuring
period A & period B, period A & A-to-B time
interval, pulse width A & A-to-B time interval,
and pulse width A & pulse width B.
IM 704510-17E