YOKOGAWA TA720 User Manual page 42

Time interval analyzer communication interface
Hide thumbs Also See for TA720:
Table of Contents

Advertisement

:CALCulation[:MEAS<x>][:WINDow<x>]:
MEDian?
Function
Queries the median value when using
histogram.
Syntax
:CALCulation[:MEAS<x>][:WINDow<x>]:
MEDian? [{POSitive|NEGative|ALL|
SELect}]
<x> of MEAS<x>=1 to 2
<x> of WINDow<x>=Window number (1
to number of windows)
Example
When in time stamp/hardware histogram
mode
• When measuring period, A-to-B time interval,
or pulse width and single window is used
:CALCULATION:MEDIAN?
-> :CALCULATION:MEDIAN 1.00145E-7
• When measuring period, A-to-B time interval,
or pulse width and multi window or auto
window is used
:CALCULATION:WINDOW1:MEDIAN?
-> :CALCULATION:WINDOW1:
MEDIAN 1.00145E-7
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B and single window is used
:CALCULATION:MEAS1:MEDIAN?
-> :CALCULATION:MEAS1:
MEDIAN 1.00145E-7
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B and multi window or auto window is
used
:CALCULATION:MEAS1:WINDOW1:
MEDIAN?
-> :CALCULATION:MEAS1:WINDOW1:
MEDIAN 1.00145E-7
When in inter-symbol interference analysis
mode
:CALCULATION:WINDOW1:MEDIAN? ALL
-> :CALCULATION:WINDOW1:
MEDIAN 1.00145E-7
Description • If the statistical value is not valid, "NAN" is
returned in response to a query.
• When using single window, omit
":WINDow<x>".
• When measuring period, A-to-B time interval,
or pulse width, omit ":MEAS<x>".
• When ":CALCulation:POLarity" is set to
"POSNeg", specify "POSitive|NEGative".
• When in inter-symbol interference analysis
mode, specify "ALL|SELect".
IM 704510-17E
:CALCulation[:MEAS<x>][:WINDow<x>]:
MINimum?
Function
Queries the minimum value when using
histogram.
Syntax
:CALCulation[:MEAS<x>][:WINDow<x>]:
MINimum? [{POSitive|NEGative|ALL|
SELect}]
<x> of MEAS<x>=1 to 2
<x> of WINDow<x>=Window number (1
to number of windows)
Example
When in time stamp/hardware histogram
mode
• When measuring period, A-to-B time interval,
or pulse width and single window is used
:CALCULATION:MINIMUM?
-> :CALCULATION:
MINIMUM 9.99437E-8
• When measuring period, A-to-B time interval,
or pulse width and multi window or auto
window is used
:CALCULATION:WINDOW1:MINIMUM?
-> :CALCULATION:WINDOW1:
MINIMUM 9.99437E-8
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B and single window is used
:CALCULATION:MEAS1:MINIMUM?
-> :CALCULATION:MEAS1:
MINIMUM 9.99437E-8
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B and multi window or auto window is
used
:CALCULATION:MEAS1:WINDOW1:
MINIMUM?
-> :CALCULATION:MEAS1:WINDOW1:
MINIMUM 9.99437E-8
When in inter-symbol interference analysis
mode
:CALCULATION:WINDOW1:MINIMUM? ALL
-> :CALCULATION:WINDOW1:
MINIMUM 9.99437E-8
Description • If the statistical value is not valid, "NAN" is
returned in response to a query.
• When using single window, omit
":WINDow<x>".
• When measuring period, A-to-B time interval,
or pulse width, omit ":MEAS<x>".
• When ":CALCulation:POLarity" is set to
"POSNeg", specify "POSitive|NEGative".
• When in inter-symbol interference analysis
mode, specify "ALL|SELect".
4.3 CALCulation Group
4-19
4

Advertisement

Table of Contents
loading