:TVARiation<x>:VERTical:SPAN
Function
Queries the span of the vertical axis (Y-axis).
Syntax
:TVARiation<x>:VERTical:SPAN
{<Time>}
:TVARiation<x>:VERTical:SPAN?
<x>=1 to 2
<Time>=500 ps, 1 ns, 2.5 ns, 5 ns,
10 ns, 20 ns, 50 ns, 100 ns, 200
ns, 500 ns, 1 µs, 2 µs, 5 µs, 10 µs,
20 µs, 50 µs, 100 µs, 200 µs, 500
µs, 1 ms, 2 ms, 5 ms, 10 ms, 20 ms
Example
:TVARIATION1:VERTICAL:SPAN 200ns
:TVARIATION1:VERTICAL:SPAN?
-> :TVARIATION1:VERTICAL:
SPAN 0.2E-06

4.22 UNIT Group

The commands in this group deal with the default unit of voltage, time, and frequency data.
:UNIT
:
:UNIT?
Function
Queries the default unit of voltage, time and
frequency.
Syntax
:UNIT?
Example
:UNIT?
-> :UNIT:FREQUENCY HZ;VOLTAGE V;
TIME S
:UNIT:VOLTage
Function
Sets the default unit of voltage or queries the
current setting.
Syntax
:UNIT:VOLTage {<Multiplier>}V
:UNIT:VOLTage?
<Multiplier> See section 3.4.
Example
:UNIT:VOLTAGE V
:UNIT:VOLTAGE?
-> :UNIT:VOLTAGE V
IM 704510-17E
;
VOLTage
<Space>
?
TIME
<Space>
?
FREQuency
<Space>
?
?
4.21 TVARiation<x> Group/4.22 UNIT Group
<Multiplier>
<Multiplier>
V
<Multiplier>
<Multiplier>
S
<Multiplier>
<Multiplier>
Hz
:UNIT:TIME
Function
Sets the default unit of time or queries the
current setting.
Syntax
:UNIT:TIME {<Multiplier>}S
:UNIT:TIME?
<Multiplier> See section 3.4.
Example
:UNIT:TIME s
:UNIT:TIME?
-> :UNIT:TIME S
:UNIT:FREQuency
Function
Sets the default unit of frequency or queries the
current setting.
Syntax
:UNIT:FREQuency {<Multiplier>}Hz
:UNIT:FREQuency?
<Multiplier> See section 3.4.
Example
:UNIT:FREQuency HZ
:UNIT:FREQUENCY?
-> :UNIT:FREQUENCY HZ
V
S
Hz
4-71