4.8 HHIStogram<x> and THIStogram<x> Group
:{HHIStogram<x>|THIStogram<x>}?
Function
Queries all settings related to the histogram
display in hardware histogram or time stamp
mode.
Syntax
:{HHIStogram<x>|THIStogram<x>}?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM?
-> :HHISTOGRAM1:MARKER:LOW 0;
STATE 1;:HHISTOGRAM1:MODE SINGLE;
SINGLE:HORIZONTAL:
CENTER 150.000E-09;SPAN 0.30E-06;:
HHISTOGRAM1:SINGLE:MARKER:
LEFT 0.0E+00;RIGHT 300.000E-09;:
HHISTOGRAM1:SINGLE:
FREQUENCY 26.315E+06;:
TVALUE 38.000E-09;:HHISTOGRAM1:
VERTICAL:AXIS LOGARITHMIC;
HIGH 0.1E+03
:{HHIStogram<x>|THIStogram<x>}:AUTO?
Function
Queries all settings related to the auto window.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
AUTO?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:AUTO?
-> :HHISTOGRAM1:AUTO:
MODULATION EFM;TTYPE ESTIMATED;
WINDOW1:STATE 1
:{HHIStogram<x>|THIStogram<x>}:AUTO:
MODulation
Function
Sets the modulation type on the auto window or
queries the current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
AUTO:MODulation {EFM|EFMPlus|M1TO7}
:{HHIStogram<x>|THIStogram<x>}:
AUTO:MODulation?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:AUTO:MODULATION EFM
:HHISTOGRAM1:AUTO:MODULATION?
-> :HHISTOGRAM1:AUTO:MODULATION EFM
4-38
:{HHIStogram<x>|THIStogram<x>}:AUTO:
TTYPe
Function
Sets how to determine constant T on the auto
window or queries the current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
AUTO:TTYPe {MEASured|ESTimated}
:{HHIStogram<x>|THIStogram<x>}:
AUTO:TTYPe?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:AUTO:TTYPE MEASURED
:HHISTOGRAM1:AUTO:TTYPe?
-> :HHISTOGRAM1:AUTO:TTYPE MEASURED
:{HHIStogram<x>|THIStogram<x>}:AUTO:
WINDow1:STATe
Function
Turns ON/OFF window 1 on the auto window or
queries the current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
AUTO:WINDow1:STATe {<Boolean>}
:{HHIStogram<x>|THIStogram<x>}:
AUTO:WINDow1:STATe?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:AUTO:WINDOW1:STATE ON
:HHISTOGRAM1:AUTO:WINDOW1:STATE?
-> :HHISTOGRAM1:AUTO:WINDOW1:
STATE ON
:{HHIStogram<x>|THIStogram<x>}:MARKer?
Function
Queries all settings related to the marker.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MARKer?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MARKER?
-> :HHISTOGRAM1:MARKER:LOW 0;
STATE 1
:{HHIStogram<x>|THIStogram<x>}:MARKer:LOW
Function
Sets the low marker value or queries the current
setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MARKer:LOW {<NRf>}
:{HHIStogram<x>|THIStogram<x>}:
MARKer:LOW?
<x>=1 to 2
<NRf>=Frequency
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MARKER:LOW 100
:HHISTOGRAM1:MARKER:LOW?
-> :HHISTOGRAM1:MARKER:LOW 100