YOKOGAWA TA720 User Manual page 62

Time interval analyzer communication interface
Hide thumbs Also See for TA720:
Table of Contents

Advertisement

:{HHIStogram<x>|THIStogram<x>}:
MARKer[:STATe]
Function
Turns ON/OFF the marker or queries the
current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MARKer[:STATe] {<Boolean>}
:{HHIStogram<x>|THIStogram<x>}:
MARKer:STATe?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MARKER:STATE ON
:HHISTOGRAM1:MARKER:STATE?
-> :HHISTOGRAM1:MARKER:STATE 1
:{HHIStogram<x>|THIStogram<x>}[:MODE]
Function
Sets the window mode or queries the current
setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}
[:MODE] {SINGle|MULTi|AUTO}
:{HHIStogram<x>|THIStogram<x>}:
MODE?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MODE SINGLE
:HHISTOGRAM1:MODE?
-> :HHISTOGRAM1:MODE SINGLE
:{HHIStogram<x>|THIStogram<x>}:MULTi?
Function
Queries all settings related to the multi window.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MULTi?
<x>=1 to 2
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MULTI?
-> :HHISTOGRAM1:MULTI:SIZE 1;
FREQEUNCY 26.315E+06;
TVALUE 38.000E-09;
OFFSET 10.000-09E;:HHISTOGRAM1:
MULTI:WINDOW1:HORIZONTAL:
CENTER 114.000E-09;SPAN 0.060E-06;:
HHISTOGRAM1:MULTI:WINDOW1:LABEL 3;
MARKER:LEFT 95.000E-09;
RIGHT 133.000E-09
IM 704510-17E
4.8 HHIStogram<x> and THIStogram<x> Group
:{HHIStogram<x>|THIStogram<x>}:MULTi:
FREQuency
Function
Sets constant T using the frequency format or
queries the current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MULTi:FREQuency {<Frequency>}
:{HHIStogram<x>|THIStogram<x>}:
MULTi:FREQuency?
<x>=1 to 2
<Frequency>=4 MHz to 1000 MHz
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MULTI:FREQUENCY 50MHz
:HHISTOGRAM1:MULTI:FREQUENCY?
-> :HHISTOGRAM1:MULTI:
FREQUENCY 50.000E+06
Description Setting/Query is made using the frequency
format of
":{HHIStogram<x>|THIStogram<x>}:
MULTi:TVALue".
:{HHIStogram<x>|THIStogram<x>}:MULTi:
OFFSet
Function
Sets the offset value on constant T or queries
the current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MULTi:OFFSet {<Time>}
:{HHIStogram<x>|THIStogram<x>}:
MULTi:OFFSet?
<x>=1 to 2
<Time>=–100 ns to 300 ns (25 ps
steps)
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MULTI:OFFSET 10ns
:HHISTOGRAM1:MULTI:OFFSET?
-> :HHISTOGRAM1:MULTI:
OFFSET 10.000E-09
:{HHIStogram<x>|THIStogram<x>}:MULTi:
SIZE
Function
Sets the number of windows or queries the
current setting.
Syntax
:{HHIStogram<x>|THIStogram<x>}:
MULTi:SIZE {<NRf>}
:{HHIStogram<x>|THIStogram<x>}:
MULTi:SIZE?
<x>=1 to 2
<NRf>=The number of windows (1 to
14)
Example
(The following is an example for the hardware
histogram mode.)
:HHISTOGRAM1:MULTI:SIZE 10
:HHISTOGRAM1:MULTI:SIZE?
-> :HHISTOGRAM1:MULTI:SIZE 10
4
4-39

Advertisement

Table of Contents
loading

Table of Contents