4.21 TVARiation<x> Group
The commands in this group deal with the time variation display in time stamp mode. You can make the same
settings and inquiries as when the DISPLAY, SCALE, and MARK keys on the front panel are used.
• When measuring the period, A-to-B time interval, or pulse width or when measuring period A & period B,
period A & A-to-B time interval, pulse width A & A-to-B time interval, and pulse width A & pulse width B, use
"TVARiation1" or "TVARiation" for the command for measurement 1. However, the query is
"TVARIATION1".
• When measuring period A & period B, period A & A-to-B time interval, pulse width A & A-to-B time interval,
and pulse width A & pulse width B, use "TVARiation2" for the command for measurement 2.
:TVARiation
<x>
:TVARiation<x>?
Function
Queries all settings related to time variation.
Syntax
:TVARiation<x>?
<x>=1 to 2
Example
:TVARIATION?
-> :TVARIATION1:HORIZONTAL:
MINIMUM 100.0E-09;SPAN 0.06E-03;:
TVARIATION1:MARKER:
HIGH 150.000E-09;LEFT 1.0E-06;
LOW 100.000E-09;RIGHT 2.0E-06;
STATE 1:TVARIATION1:VERTICAL:
CENTER 100.000E-09;SPAN 0.2E-06
IM 704510-17E
;
:
HORizontal
:
MARKer
:
VERTical
:
4.21 TVARiation<x> Group
;
MINimum
<Space>
?
SPAN
<Space>
?
?
;
STAT e
<Space>
HIGH
<Space>
LOW
LEFT
<Space>
RIGHt
LV ALue
?
RVALue
?
;
CENTer
<Space>
SPAN
<Space>
:TVARiation<x>:HORizontal?
Function
Queries all settings related to the horizontal axis
(X-axis) of the time variation display.
Syntax
:TVARiation<x>:HORizontal?
<x>=1 to 2
Example
:TVARIATION:HORIZONTAL?
-> :TVARIATION1:HORIZONTAL:
MINIMUM 100.0E-09;SPAN 0.06E-03
<Time>
<Time>
OFF
ON
<NRf>
?
<Time>
?
<Time>
?
<Time>
?
<Time>
?
?
?
4-69