4.14 SAMPle Group/4.15 SSTart Group/4.16 STARt Group
:SAMPle:GATE[:MODE]
Function
Sets the gate type or queries the current setting.
Syntax
:SAMPLe:GATE[:MODE]
{EVENt|TIME|EXTernal}
:SAMPle:GATE:MODE?
Example
:SAMPLE:GATE:MODE EVENT
:SAMPLE:GATE:MODE?
-> :SAMPLE:GATE:MODE EVENT
:SAMPle:GATE:POLarity
Function
Sets the polarity of the external gate or queries
the current setting.
Syntax
:SAMPLe:GATE:POLarity
{POSitive|NEGative}
:SAMPle:GATE:POLarity?
Example
:SAMPLE:GATE:POLARITY POSITIVE
:SAMPLE:GATE:POLARITY?
-> :SAMPLE:GATE:POLARITY POSITIVE
Description Setting is possible when "SAMPle:GATE:MODE"
is set to "EXTernal".
:SAMPle:GATE:TIME
Function
Sets the gate time in terms of time or queries
the current setting.
Syntax
:SAMPLe:GATE:TIME {<Time>}
:SAMPle:GATE:TIME?
<Time>=1 µs to 10 s (100 ns steps)
Example
:SAMPLE:GATE:TIME 1us
:SAMPLE:GATE:TIME?
-> :SAMPLE:GATE:TIME 1.0E-06
Description Setting is possible when "SAMPle:GATE:MODE"
is set to "TIME".

4.15 SSTart Group

:SSTar t
:SSTart
Function
Executes single measurement.
Syntax
:SSTart
Example
:SSTART

4.16 STARt Group

:STARt
:STARt
Function
Starts the measurement.
Syntax
:STARt
Example
:START
4-64
:SAMPle:INHibit
Function
Sets the polarity of inhibit input or queries the
current setting.
Syntax
:SAMPLe:INHibit
{OFF|POSitive|NEGative}
:SAMPle:INHibit?
Example
:SAMPLE:INHIBIT OFF
:SAMPLE:INHIBIT?
-> :SAMPLE:INHIBIT OFF
Description When set to "OFF", the inhibit input is invalid.
:SAMPle:INTerval
Function
Sets the sampling interval or queries the current
setting.
Syntax
:SAMPLe:INTerval {<Time>|MINimum}
:SAMPle:INTerval?
<Time>=0 to 1 s (1 µs steps)
Example
:SAMPLE:INTERVAL MINIMUM
:SAMPLE:INTERVAL?
-> :SAMPLE:INTERVAL MINIMUM
Description • If "MINimum" is transmitted for the data or 0 is
transmitted for <Time>, sampling is
performed using the minimum interval.
• Setting is possible when the sampling mode
is set to time stamp.
:SAMPle:RCLock
Function
Sets reference sampling clock or queries the
current setting.
Syntax
:SAMPLe:RCLock {INTernal|EXTernal}
:SAMPle:RCLock?
Example
:SAMPLE:RCLOCK INTERNAL
:SAMPLE:RCLOCK?
-> :SAMPLE:RCLOCK INTERNAL