YOKOGAWA TA720 User Manual page 38

Time interval analyzer communication interface
Hide thumbs Also See for TA720:
Table of Contents

Advertisement

:CALCulation[:MEAS<x>][:{BLOCk<x>|
BALL}]:TSNumber?
Function
Queries the number of samples for the
statistical computation when using time
variation.
Syntax
:CALCulation[:MEAS<x>][:{BLOCk<x>|
BALL}]:TSNumber?
<x> of MEAS<x>=1 to 2
<x> of BLOCk<x>=Block number (1 to
number of blocks)
BALL: Queries the statistical value
of the entire block.
Example
• When measuring period, A-to-B time interval,
or pulse width and block sampling is not used
:CALCULATION:TSNUMBER?
-> :CALCULATION:TSNUMBER 1.000E+3
• When measuring period, A-to-B time interval,
or pulse width and block sampling is used
:CALCULATION:BLOCK1:TSNUMBER?
-> :CALCULATION:BLOCK1:
TSNUMBER 1.000E+3
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B
:CALCULATION:MEAS1:TSNUMBER?
-> :CALCULATION:MEAS1:
TSNUMBER 1.000E+3
Description • If the statistical value is not valid, "NAN" is
returned in response to a query.
• When the computation range is a block and
block sampling is used, specify
":{BLOCk<x>|BALL}".
• When measuring period, A-to-B time interval,
or pulse width, omit ":MEAS<x>".
IM 704510-17E
:CALCulation[:MEAS<x>]:CONStt
Function
Sets constant T or queries the current setting.
Syntax
:CALCulation[:MEAS<x>]:CONStt
{<Time>}
:CALCulation[:MEAS<x>]:CONStt?
<x>=1 to 2
<Time>=1 ns to 250 ns (25 ps steps)
Example
• When measuring period, A-to-B time interval,
or pulse width
:CALCULATION:CONSTT 100ns
:CALCULATION:CONSTT?
-> :CALCULATION:CONSTT 100.000E-9
• When measuring period A & period B, period
A & A-to-B time interval, pulse width A & A-to-
B time interval, and pulse width A & pulse
width B
:CALCULATION:MEAS1:CONSTT 100ns
:CALCULATION:MEAS1:CONSTT?
-> :CALCULATION:MEAS1:
CONSTT 100.000E-9
Description • Setting and query are possible when single
window or multi window is enabled.
• When measuring period, A-to-B time interval,
or pulse width, omit ":MEAS<x>".
4.3 CALCulation Group