5.0 Operation
Fig. 29. If the foot regulator is moved towards the toe
stop, the control of the vacuum passes to the
foot regulator at about -34kPa
Fig. 30. Display after control of the vacuum has
passed to the foot regulator
Fig. 31. If the foot regulator is moved towards the
heel stop, the control of the vacuum passes
to the foot regulator at about -52kPa
24
5.1.6 Use of the foot regulator in VE
Mode
During VE mode, it is possible to transfer the control of the
vacuum to the connected foot regulator. The ATMOS S 351
Natal automatically goes into Suction mode in this case.
The ATMOS S 351 Natal must be switched off before the foot
regulator is connected or disconnected.
In VE mode, set the foot regulator pedal towards the positi-
on which corresponds to the currently applied vacuum (fi g.
29 and 31). This transfers control over the vacuum to the
foot regulator. The ATMOS S 351 Natal goes into suction
mode (fi g. 30). The automatic vacuum build up mode can
be restarted by pressing the extraction cup button (fi g.
19).