5.0 Operation
Fig. 15. Display directly after switching on
18
Function vacuum extraction
For vacuum extraction, we recommend to use a small coll-
ection jar (1.5 l) to enable a quick generation of the required
vacuum value.
Automatic vacuum build up
During automatic vacuum build up, the vacuum is gener-ated
evenly and fi nal vacuum is reached in the preset time (standard
adjustment = 120s). For changing this vacuum build-up time,
refer to chapter 5.3 Adjustments, page 28. During vacuum
generation, the green LED fl ashes. When the fi nal vacuum
is reached the device gives an audible signal and the green
LED lights up constantly.
The device offers the possibility to select between a fully
automatic vacuum build-up (VE Auto Mode) and a semi-
automatic vacuum build-up (VE Semi Mode) (see chapter 5.3
Adjustments, page 29).
The fully automatic vacuum build-up is standardly pre-
set.

5.1.2 Fully automatic vacuum build up

(VE Auto Mode)

Switch on the ATMOS S 351 Natal. Make sure that the pilot
lamp in the switch lights up.
Then, the switching-on menu (fi g. 15) is shown in the
display for approx. 2seconds.