Faults; Emergency Operation; Overheating; Fault - Atmos S 351 Operating Instructions Manual

Hide thumbs Also See for S 351:
Table of Contents

Advertisement

5.0 Operation
Fig. 44. Display for emergency operation
Fig. 46. Overheating
Fig. 47. Call the ATMOS service department
34
5.6

Faults

5.6.1

Emergency operation

This operating mode is automatically entered if there is a
malfunction in the ATMOS S 351 Natal. Please call the service
department if this happens. The ATMOS S 351 Natal functions
are restricted in this case. Control of the vacuum is no longer
possible. The following functions are still available:
Fig. 45.
Switch on pump
Switch on pump
Switch off pump
5.6.2

Overheating

The ATMOS S 351 Natal is designed for continuous operation.
If the ventilation louvres at the bottom are blocked, the unit
may overheat. Initially, a warning is given (graphics display
blinks):
Check the ventilation louvres of the ATMOS S 351 Natal.
If the temperature continues to rise, the ATMOS S 351 Natal
will be switched off. If this happens, the overheating display
is left on the graphics display. Please call the service depart-
ment if this happens.
Set up the ATMOS S 351 Natal on a clear, level surface.
This will ensure free air circulation.
5.6.3

Fault

If the ATMOS S 351 Natal detects a functional fault, this will
be shown on the graphics display. Please call the service
department if this happens.
Please note that the Key Error is also created when a key
is pressed for 7 seconds (exception: -keys).