Display symbols

Figure
4.3

Explanation of the display in key lock modes

4.3.1

Key lock mode with bubbles

In the key lock mode the flow is additionally displayed in traffic light coloured bubbles for at least
one hour.
4.3.2

Key lock mode with bar chart

If the key lock is active and the average flow value is less than 450 ml/min for at least one hour,
the therapy process is displayed in a bar chart over 24 hours. The flow is visualised as an average
value in bars on an hourly basis. The vacuum is displayed as a gradient line.
28
Battery status indicator / charging indicator
Key lock activated
Upcoming warning suppressed
Annual inspection is required
Key lock acti-
vated
Flow displayed as bubbles
Each additional coloured
bubble represents an
additional flow.
None: 0 - < 50 ml/min
Green: 50 - < 100 ml/min
Yellow: 100 - < 630 ml/min
Orange: 630 ml - < 2.01 l/
min
Red: >2.01 l/min to
maximum
Up to 1.00 l/min the flow is
displayed in ml/min.
The bar chart appears with a max. flow scaling of 450
ml/min if the average flow value is less than 450 ml/min
for at least one hour.
Function
Day/Night Mode
The ATMOS C 051 Thorax has a day/
night mode, i.e. the device adjusts
automatically to the light conditions
in a room.
Under low ambient light conditions,
the display has dark background
illumination.