5.0 Operation
26
Presetting the vacuum with the foot regulator
The ATMOS S 351 Natal must be switched off before the foot
regulator is connected or disconnected. Completely depress
the connected foot regulator to the heel position.
Switch on the ATMOS S 351 Natal. Check that the lamp in
the switch is illuminated.
After the VE menu is displayed, switch into suction mode
by pressing the END button (or the
Using the
button, select the vacuum value required which
is to be built up with the foot regulator (see vacuum / time
bar graph).
Apply the extraction cup and build up the vacuum required
in steps using the foot regulator (the foot regulator remains
in position if you take your foot away).
Switch into VE mode by pressing the extraction cup button.
The ATMOS S 351 Natal starts automatic vacuum build up
from the vacuum value set in suction mode. The time until
fi nal vacuum is shown at the top left of the display.
button).