Additional Functions - Atmos S 351 Operating Instructions Manual

Hide thumbs Also See for S 351:
Table of Contents

Advertisement

5.0 Operation
Fig. 28. Buttons for increasing /
decreasing the vacuum

5.1.5 Additional functions

In the VE mode of the ATMOS S 351 Natal it is possible at any
time to intervene in the operation of the unit using the arrow
buttons (fi g. 28). The Max button is locked.
Using the   buttons
Operation of the arrow buttons in VE mode is signalled by
an alert tone. The buttons have the following functions in this
mode:
The vacuum build up can be delayed by pressing the
button during vacuum build up. The counter for the
time indication is increased accordingly. You can leave the
VE mode by pressing the
The vacuum build up can be accelerated by pressing the
button during vacuum build up. The counter for the time
indication is decreased accordingly.
Once the fi nal vacuum has been reached, it can be reduced
by pressing the
button. You can leave the VE mode by
pressing the
button several times.
Once the fi nal vacuum has been reached, it can be in-
creased by pressing the
The vacuum release can be accelerated by pressing the
button during vacuum release. The counter for time
indication is decreased accordingly. You can leave the
VE mode by pressing the
The vacuum release can be delayed by pressing the
button during vacuum release.
The counter for time indication is increased accordingly.
button several times.
button.
button several times.
23