5.0 Operation

5.1.7 Vacuum build up with foot regulator

The ATMOS S 351 Natal must be switched off before
the foot regulator is connected or disconnected.
Completely depress the connected foot regulator to the
heel position.
Switch on the ATMOS S 351 Natal. Check that the indicator
lamp in the switch is illuminated.
After the VE menu is displayed, switch over into suction
mode by pressing the END button (or the
Using the
button, select the required vacuum value
which is to be built up with the foot regulator (see vacuum /
time bar graph).
Apply the extraction cup and build up the fi nal vacuum
required in steps using the foot regulator (the foot
regulator remains in position if you take your foot away).
5.1.8 Starting automatic vacuum build up
from a vacuum value preset in suction
mode
It is possible for the ATMOS S 351 Natal to start the automa-
tic vacuum build up from a vacuum value preset in suction
mode:
Manual presetting of vacuum
Switch on the ATMOS S 351 Natal. Check that the indicator
lamp in the switch is illuminated.
After the VE menu is displayed, switch over into suction
mode by pressing the END button (or the
Apply the extraction cup and manually build up the vacuum
in steps to the level required (by pressing the
Change into VE mode by pressing the extraction cup sym-
bol. The ATMOS S 351 Natal starts automatic vacuum build
up from the vacuum value set in suction mode. The time
until fi nal vacuum is shown at the top left of the display.
button).
button).
button).