Display
A safety check (inspection according to the manufacturers specifi cations)
is required every year. This will be displayed on the device.
Low battery capacity appears in the display. Battery must be replaced by
the service.
If the target vacuum is adjusted over-50 mbar a notice appears
,,High target vacuum is set"

6.0 Functions

6.1 Hose rinsing

The ATMOS
C 051 Thorax has an automatic hose rinsing function which works periodically.
®
The rinsing process transports secretion located in the secretion hose into the secretion canister.
The rinsing is performed by opening a valve located in the measuring
The manufacturers default setting for the period between 2 rinsing cycles is 5 minutes.
Cause of error
If the therapy hasn`t started after
initial operation the error indication
for "inactive therapy" is displayed.
Troubleshooting
Start therapy by pressing the
button.
channel.
31