Pressure Test
For running a pressure test, please refer to the corresponding pump manual.
The pressure test may be used for testing the tightness of a valve installed in
the TCC.
Wrong use of pressure test may damage valve.
C A U T I O N
The current implementation of the pressure test automatically uses the maximum
pressure generated by the pump used by that system.
➔
1200 Series TCC User Manual
Do not use the test for modules having a lower maximum pressure than the pump
as this will damage the valve. For example do not use 400 bar valve in a TCC in
combination with a 600 bar pump.
6
Test Functions
Pressure Test
65