Connect The Column To The Inlet And Quickswap; Configure The Column - Agilent Technologies QuickSwap G3185B Installation And Setup

Table of Contents

Advertisement

Overview

10. Connect the Column to the Inlet and QuickSwap

Inlet end
See the instructions for the inlet you are using in your GC documentation.
Much of this manual assumes that you will use a split/splitless inlet. This is
the most popular choice.
QuickSwap end
1 Insert the end of the column with the swaged ferrule into Fitting 2 on
2 Gently push until the column end bottoms out. Follow with the nut.
3 Finger-tighten the nut. Use a wrench to tighten the nut about 15 degrees
If the ferrule is not over-tightened when the column is installed in QuickSwap,
it should provide an air-tight and leak-free seal through many removals and
re-installations.

11. Configure the Column

See your MSD ChemStation manuals for the details.
The tasks to be performed are:
The Outlet psig that you enter must be the same as the purge gas supply pressure for
C A U T I O N
accurate display and control of column flow rate during the run. For checkout methods,
this must be 4.0 psig. This is also good for most purposes. Whenever you change either
of these settings, you must change the other to match or the displayed column flow
will be inaccurate.
28
QuickSwap (see
Figure 10
more.
a Define the column dimensions (required so that the GC can calculate
flows and pressures).
b Select a column flow mode (usually constant pressure or constant flow).
c Set the column outlet pressure. Under the Columns icon, select Detector
and Other. Select Positive and enter an Outlet psig.
on page 25).
QuickSwap Installation and Startup