Instructions for use SMARTmatic S80 K - 1.011.6771
7 Reprocessing steps in accordance with ISO 17664 | 7.3 Disinfection

7.3.1 Manual external disinfection

KaVo recommends the following products based on compatibility of the materials.
The microbiological efficacy must be ensured by the disinfectant manufacturer and
proven by an expert opinion.
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
Cloths for wiping the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow
the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶ Follow the instructions for use of the disinfectant.

7.3.2 Manual internal disinfection

The efficacy of manual internal disinfection must be demonstrated by the manufac-
turer of the disinfection agent. With KaVo products, use only disinfection agents that
have been released by KaVo with respect to the compatibility of materials (e.g. WL-
cid / made by ALPRO).
▶ Immediately after internal disinfection, lubricate the KaVo medical device with care
agents from the KaVo care system.

7.3.3 Automated external and internal disinfection

KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
▶ Immediately after drying, lubricate the medical device with care agents from the
KaVo care system.
The drying procedure is normally part of the cleaning program of the thermodisin-
fector.
Note
Please observe the instructions for use of the thermodisinfector.