▶ Connect the type 4882 power supply to the socket of the ELECTROmatic.
▶ Connect the power cable first to the power supply and then to the supply
mains socket.
▶ Route the cables appropriately such that there is no kinking and, if possible
and permitted by the treatment unit, in the support arm.
▶ Affix the cord with cord ties and/or cord tape.

5.2 Calibrating the foot control

▶ Press the foot control down once as far as it will go (maximum pressure) in
order to calibrate the foot control.
ð
This starts up the motor and the system calibrates automatically to the ex-
isting system pressure.
Automatic calibration of the foot control
The calibration is performed automatically during the first startup, when the
foot control is pressed for the first time.
Once the calibration of the foot control has been done once, the calibration to
the maximum system pressure takes place automatically in the background
during operation (automatic calibration).
Minor pressure fluctuations are balanced automatically in this process.