Oldest Report Id; Newest Report Id; Selected Protocol-Id; Protocol - Siemens SIWAREX WP251 Manual

Electronic weighing system
Table of Contents

Advertisement

Scale parameters and functions
8.36 DR 46 Protocol content
Variable
Note
String header for
String header for time
time
Date, time
Time
(Page 159)
String header for
String header for addi-
additional string
tional string
Additional string
Currently not used
Reserve
Reserve
Reserve
Reserve
8.36.2

Oldest report ID

The ID of the first saved log is displayed here.
8.36.3

Newest report ID

The ID of the last saved log is displayed here.
8.36.4

Selected protocol-ID

The ID of the log requested in data record 45 and shown in data record 46 is displayed here.
8.36.5

Protocol

The 46 bytes of the log have the following structure:
Designation
Gross/Net
SEMICOLON
G/N weight calibratable
SEMICOLON
Weight unit
SEMICOLON
Tare sign
SEMICOLON
Tare
SEMICOLON
158
Type
Length
Rw Default
(bytes)
UBYTE[2]
2
r
CHAR[8]
8
r
UBYTE[2]
2
r
CHAR[4]
4
r
USHORT
2
r
FLOAT
4
r
Data type
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
UBYTE
Min.
Max.
8,8
"23:59:59"
2,2
" "
0
0
Manual, 12/2015, A5E37203357A
Modbus
regis-
ters
4380
4381
4385
4386
4388
4390
Length
2
1
8
1
4
1
2
1
8
1
SIWAREX WP251

Advertisement

Table of Contents
loading