If the standstill waiting time is equal to zero, a command requiring standstill is rejected
immediately if standstill is not achieved at the time the command is issued.
If the standstill waiting time is active, this is indicated in the NAWI Status of the scale by a
corresponding bit.
8.3.25

Standstill 2 - range

This parameter defines the weight range (+/- around the current weight) by which the weight
may fluctuate within Standstill time 2, in order to determine that standstill has occurred. The
setting is made in numerical increments (d).
8.3.26

Standstill 2 - time

This parameter defines the time window in which the weight value may fluctuate within
Standstill range 2 only, in order to determine that standstill has occurred. The setting is made
in milliseconds (ms).
8.3.27
Settling time before standstill 2
The parameter defines a time in milliseconds (ms) that is allowed to elapse before a check is
made for standstill 2. Because standstill 2 is needed after an automatic dosing for logging or
for the tolerance check, the filled scale, which may not yet be at steady state, can settle
before the check for standstill 2 is made. If the settling time is active, this is indicated in the
AWI Status of the scale by a corresponding bit.
SIWAREX WP251
Manual, 12/2015, A5E37203357A
Scale parameters and functions
8.3 DR 3 Calibration parameters