Siemens SIWAREX WP251 Manual page 202

Electronic weighing system
Table of Contents

Advertisement

Operation requiring official calibration
12.2 Verification
8. Checking of the additive tare range
When using the additive tare compensation unit, you must check the complete weighing
range (up to maximum load + additive maximum tare load). To do this, you must check
up to the maximum load, and repeat this following taring. Repeat these steps until the
upper limit of the range of the additive tare compensation unit has been reached.
– Call up the main menu in the main display (function key with the open-end wrench).
– Select the "Scale check" submenu in the "Calibration check" submenu.
– Check with the "Tare setting" or "Delete tare" command.
9. Checking of the parameter disabling/calibration bridge
The calibration bridge must be inserted on the weighing module. You can check the
calibration bridge using the specific symbol
(symbol displayed = calibration bridge inserted).
10.When using the second display S102 - checking of set telegram address
The telegram address (No. 01) must be set on the Siebert display S102. Checking is
carried out in accordance with the specifications in the Instruction Manual "Siebert series
S102 - digital displays for Siemens Siwarex".
202
Select the "Calibration check" submenu in the main menu.
at the bottom right in the main display
Manual, 12/2015, A5E37203357A
SIWAREX WP251

Advertisement

Table of Contents
loading

Table of Contents