2.0 Front Panel Button Test

Tools Needed: QED 6 Defibrillator Analyzer
Test Setup:
24.
Install strip chart paper into the recorder tray.
25.
Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.
Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the ZOLL simulator, or Fluke Biomedical Impulse 4000 Analyzer (or
26.
equivalent).
Do this...
2.1
Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
2.2
Set the QED 6 analyzer to NSR of 120 BPM. To
check the size of the ECG waveform, press the
SIZE button.
2.3
Press the ALARM SUSPEND button.
2.4
Press the RECORDER button.
2.5
Open the paper compartment door, then press
the RECORDER button.
2.6
Close the paper compartment door, then press
the RECORDER button.
2.7
Press the RECORDER button.
2.8
Press the VOLUME softkey, then press Prompt
Volume.
9650-1450-01 Rev. D
Verify that...
The unit sounds 4 beep tones. PADS and MONITOR (non-AED units) display
on the monitor.
Note: PADS is a factory default setting.
As you press the SIZE button five times (0.5, 1.0, 1.5, 2.0, 3.0), the size of the
ECG waveform appropriately changes on the display.
Alarm symbol changes from disabled to enabled. If the alarm sounds, press
the ALARM SUSPEND button to turn it off. The alarm will only be suspended
for 90 seconds at this point. Press and hold the ALARM SUSPEND button for
3 seconds to disable alarms.
The strip chart paper moves out of the unit from the paper tray. Check that the
correct time, date, ECG lead annotation and waveform are recorded on the
paper. (Set Time and Date, if necessary.)
The CHECK RECORDER message appears on the monitor.
The strip chart paper flows out of paper tray and the monitor no longer displays
the CHECK RECORDER message.
The strip chart paper stops flowing out of the paper tray.
The volume bar graph displays.
Note: The QRS tone is on or off. There is no gradual change in volume. If
equipped, voice prompts are gradual. Note: The voice volume has 5
settings. Setting 3 is in the mid-range.
E Series Service Manual
Pass/Fail/NA
o o
o o o
o o o
o o o
o o o
o o o
o o o
o o
1-7