(1)-5 Misc.
Setting
(1)
Span
(2)
Annotation
(3)
Perform Auto Zero ADJ
(4)
Reset Auto Zero ADJ
(5)
[Zero Position Voltage Value]
Logic and Pulse settings
Specifies the digital input settings.
(1)-6 Logic/Pulse
Selects the digital input processing method.
Setting
Off
Disables the measurement of digital input.
Logic
Digital input is processed as logic signals.
Pulse
Digital input is processed as pulse signals.
(1)-7 Input
Sets the pulse measurement mode. This item can be set only when Pulse is selected in (1)-6.
Setting
Off
Disables the pulse input.
Revol.
Counts the number of pulses per second and and multiplies it by 60 to capture an rpm value.
Counts
Captures the cumulative number of pulses for each sampling interval from the start of measurement.
Inst.
Captures the number of pulses for each sampling interval.
(1)-8 Filter
Sets the filter for digital input.
Setting
Off
Disables the hardware filter.
On
Enables the hardware filter. The filter is effective in an environment where there is lots of noise.
Filter is approximately 30 Hz (-3dB).
Sets the upper and lower limits of the span in which waveforms should be
displayed.
Sets an annotation (comment) to be displayed for each channel.
The annotation can be 11 characters long at the maximum.
You can use alphanumeric and kana characters and symbols.
(For details on text input, see page 3-41.)
Performs calculation using the current input voltage as the zero position
voltage value.
Cancels the zero position voltage value and displays the input voltage.
Displays the zero position voltage value (display only).
Description
Description
Description
Settings and Measurement
Description