Settings and Measurement
(1)-9 Slope
Sets the slope (direction) to count the number of pulses. This item can be set only when Pulse is
selected in (1)-6.
Setting
↑ H
↓ L
(1)-10 EU (Scaling)
Converts the measured signals to other units.
(1)
Function
(2)
Meas. Value
(Upper and Lower Limits)
(3)
EU Output Value
(Upper and Lower Limits)
(4)
Dec pt
(5)
Select
(6)
Unit
(7)
User-Defined Unit
CHECKPOINT
• If a message appears, follow the instructions by reducing the number of digits to be output by one,
or leaving the number of digits as is and changing the EU value.
• The Scaling operation is calculated using a ratio of the Meas. Value or EU Output Value settings.
"++++/----" is displayed when the converted value cannot be processed by MT100.
Span may be changed according to the value set for Scaling.
Setting Example
Upper Value
Lower Value
+5 V
CH.1 10V
–5 V
3-22
Description
Counts the rising edges of pulses.
Counts the falling edges of pulses.
Setting
Sets On/Off of the Scaling function.
Sets the upper and lower limits of the numeric value to be converted.
Sets the upper and lower limits of the output value after conversion.
Specifies the decimal point position of the EU output value(s).
Selects the specific type of engineering units (The following selections are
available).
Current, length, area, volume, speed, acceleration, frequency, weight, power,
pressure, flow, temperature
Selects the converted unit.
The units displayed in this item are those of the type selected in "Select."
To set a unit not displayed in this item, define an arbitrary character string in "User-
Defined Unit." The setting made in this item is displayed in "User-Defined Unit"
Selects the converted unit, which can be specified as a user-defined character string
consisting of alphanumerics.
(For details on text input, see page 3-41.)
The settings made in "Select" and "Unit" are reflected here.
Meas. Value
+ 5.000
–5.000
Description
EU Value
Dec pt
+ 20.00
+ xx.xx
–20.00
+20.00 rpm
CH.1 Scaling 1
–20.00 rpm
Unit
rpm